EMFC MK III E.L.F. Magnetic Field Compensation

EMFC MK III E.L.F. Magnetic Field Compensation

To maximize performance:

  • Make sure detector is aligned with the axes of the room (x,y,z)
  • Disable all channels. Close door of BOTH EM rooms. Turn off video
    monitors. Enable all channels.
  • Plug BNC cable into X channel and observe on oscilloscope. Use small
    screwdriver to increase gain of X channel until feeback observed on
    oscilloscope. Reduce gain by 0.5 turns.
  • Repeat same procedure for Y channel and Z channel.
  • Do this sequence several times until system converges.